The AX-80 mass spectrometer system can be configured to meet most
Residual gas
analysis and process monitoring applications. With mass ranges of 80amu and a choice of detector systems the
AX-80 may be supplied as an
"add on" to an existing vacuum system or as a complete stand
alone unit with integrated vacuum and inlet system. The AX-80 uses a PC
Computer as its system controller and its operating software is written in
a high level language. This simplifies updating, and allows easy
customising to meet specific process requirements. The user-friendly
software operating under 'Windows' make the AX-80 easy to use in all RGA
and gas analysis applications.
Key Benefits
 | Low Cost |
 | Small Size |
 | Simple to use |
 | Comprehensive Software. |
Complete RGA.
|
The Analyser.>
The quadrupole analyser is available with a choice of ion sources,
mass filters and detection systems. The standard ion source is a
radially symmetrical design fitted with twin filaments and suits most
applications. The standard filaments are tungsten but thoria coated
iridium are available if required.
|
waiting for picture
|
|
^The Electronics Unit.
This compact unit contains the complete quadrupole electronics and
plugs directly onto the analyser. Communication with the PC Computer is
via the RS232.
The unit has a performance and features only found on more expensive
RGA's, and yet its small size and low cost is achieved by the use of the
very latest electronic techniques. |
|
Software features
The new 'Windows' RGA software for the AX-80 mass spectrometer has
been designed to be 'user friendly' and very simple to learn and use.
For detailed information refer to the separate software
leaflet.
|
|
Technical Data.
|
|
Mass Range
|
1-80
|
|
Resolution
|
10% valley between equal peaks.
|
|
Minimum detectable partial pressure
|
1x10-10mBar Faraday,
|
|
Minimum detectable total pressure
|
5 x 10 -11
mBar.
|
|
Maximum operating pressure
|
1x10-2mBar Faraday,
|
|
Electron energy
|
80eV
|
|
Ion Energy
|
6eV
|
|
Emission
|
100 uA *
|
|
Bake-out temperature
|
250C recommended. 350C maximum
|
|
R.F frequency
|
1.84Mhz std.(options 1.6Mhz - 4.9Mhz)
|
|
R.F. amplitude
|
800V p/p at maximum mass
|
|
Rod diameter
|
3.0 m.m.
|
|
Rod length
|
20 m.m.
|
|
Power requirements
|
17-26VDC 15Watt
|
|
Software specification
|
Refer to separate Software
Leaflet.
|
|
P.C .Computer required.
|
Pentium preferred.
|
|
Order Information. - AX-80.
|
|
|
Complete Instruments
|
AX-80/80F
|
80amu RGA with Faraday Detector
|
|
Spare Analyzer Heads
|
AH80F
|
80amu Analyzer with Faraday Detector
|
|
Spare Filaments
|
SM100FT
|
Pack of 2 Thoria coated Iridium
|
|
Dimensions
Electronics unit 6 x 3 x 3.5 inches (152 x 76 x 88mm)
|

|